Endress+Hauser
| Measuring Principle : Conductive, coaxial electrodes | Application : Pure and ultrapure water in life sciences and food |
| Measuring Range : 0.04 to 500 µS/cm | Brand : Endress+Hauser |
File Download
| MEASURING PRINCIPLE |
| Measuring Principle : Conductive, with electropolished stainless steel coaxial electrodes |
| APPLICATION |
| Main Application : Conductivity measurement in pure and ultrapure water in life sciences and food |
| Application : WFI |
| Application : Electro-deionization |
| Application : Ion exchangers |
| Application : Reverse osmosis |
| Application : Distillation |
| MEASURING RANGE / CONNECTIONS |
| Measuring Range : 0.04 to 500 µS/cm (k = 0.1) |
| Process Connections : Varivent, Neumo BioControl, Clamp |
| PROCESS CONDITIONS |
| Process Temperature : -5 to 120 °C (23 to 248 °F) |
| Sterilization Temperature : Max. 150 °C at 5 bar (302 °F at 72.5 psi) |
| Pressure : 13 bar at 20 °C (188 psi at 68 °F) |
| Pressure : 9 bar at 120 °C (130 psi at 248 °F) |
| WETTED MATERIALS |
| Electrodes / Process Connection : Electropolished stainless steel 1.4435 (316L) |
| Insulation : PEEK |
| Seal : ISOLAST (FFKM) or EPDM (options) |
| KEY FEATURES |
| • Sterilizable and autoclavable with replaceable seal |
| • Non-contact inductive signal transmission via Memosens 2.0 |
| • Extended data storage for predictive maintenance |
| Brand : Endress+Hauser |
| Product Family : Memosens |